HomeProductsAED & DefibrillatorsDefibrillator Monitor (Lifegai...

Defibrillator Monitor (Lifegain CU-HD1)

R190,885.49 Inc. Vat

Add to wishlist
Share

    Defibrillator Monitor (Lifegain CU-HD1)

    1. Product overview Defibrillator Monitor (Lifegain CU-HD)

    LifeGain CU-HD1 – this is the company’s high end defibrillator/monitor intended for hospital use. It has defibrillation capabilities in AED and manual modes and may use disposable electrode pads and external and internal paddles for shock delivery. It has a transcutaneous pacer for temporary pacing of patients with low heart rate. It has an SPO2 module for the monitoring of oxygen saturation in the patient’s blood and a 12-lead ECG monitor for the evaluation and monitoring of the patient’s ECG.

    It has a memory card port for the storage of acquired signals and an integrated printer for the generation of hard copies of the patient’s physiological signals. These features make this defibrillator suitable for use inside a hospital.This product is a defibrillator designed for qualified emergency staff and medical professionals to deliver defibrillating shocks to victims of sudden cardiac arrest and restore the normal ECG rhythm.

    * AED Mode

    AED Mode is delivered to patients who are exhibiting the symptoms of Sudden Cardiac Arrest (SCA) including ventricular fibrillation, ventricular tachycardia, in order to bring them back to life by delivering electric shock and restoring the normal ECG rhythm.

    In the AED mode, defibrillation pads must be used. These defibrillation pads work to obtain the ECG signal of patients and deliver electric shock according to the patients’ condition.

    * Manual Mode

    Manual Mode is divided into two functions such as asynchronous defibrillation and synchronous cardioversion. In asynchronous defibrillation, the usage target is the same in AED Mode.

    For the asynchronous defibrillation treatment in the manual mode, the user is able to choose the level of electric shock energy from the range of 1-200 joule by the use of pads or paddles.

    On the other side, when used for synchronous cardioversion, it is used on patients with the symptoms of atrial fibrillation. It is designed to analyze the R-wave of ECG QRS and deliver R-wave synchronized electric shocks.

    In the Manual mode, synchronous cardioversion treatment can be delivered to patients with rapid atrial fibrillation, ventricular tachycardia, and cardiac ischemia.

    * Pacer Mode

    Pacing is a method applied to patients who had lost natural cardiac movement functions, mostly used on patients with bradycardia. LiFEGAIN CU-HD1 functions to support non-invasive pacing, a way of helping maintain a patient’s pulse by attaching its electrode to the patient’s skin and delivering artificial electric stimulation to the heart. Pacing mode is divided into the ‘Fixed mode’ and the ‘Demand mode’.

    * Monitoring Mode

    Patient monitoring mode features the ECG monitoring function and the function to measure the level of SpO2, functional oxygen saturation in the blood.

    For the ECG monitoring function, you can discriminatingly use the 3-lead, 5-lead, 12-lead ECG cables. During the patient monitoring session, it is possible to analyze the ECG results to use the alarming function according to the conditions such as the number of pulses, ventricular fibrillation or ventricular tachycardia etc. If using an ECG cable, convert “Rotary Switch” to “Monitor mode” before usage.

    * Key features

    • Manual and AED Operation
    • Defibrillation using paddles or pads
    • Synchronized cardioversion
    • Efficient and effective e-cube Biphasic technology (BTE Type)
    • SpO2 pulse oximetry with alarm
    • ECG monitoring (3 Lead ECG / 5 Lead ECG / 10 Lead ECG)
    • Noninvasive pacing mode

    Specification

    Model Name
    Item Description
    Lifegain CU-HD1
    Defibrillator / Monitor
    Parameter Description
    Specified
    DEFIBRILLATOR
    Operating Modes Manual / Semi automatic
    Waveform Biphasic (Truncated exponential type)
    Shock Delivery Via Paddles or multi-function defibrillator electrode pads
    Reusable External Paddle Yes
    Pediatric Paddle Yes
    Multifunction Defi Pads (Disposable) Yes
    Energy sequence Yes
    Manual mode, J 1~10, 15, 20, 30, 50, 70, 100, 120, 150, 170, 200 J
    Synchronous Cardioversion Yes
    AED mode, J Fixed energy at 200 Joules
    Text and Voice Prompts Yes
    Protocol configured Yes
    ECG MONITOR
    ECG acquisition Disposable electrodes
    Monitors with ECG YES
    Input Lead I, II, III (3 Lead ECG cable)
    Lead I, II, III, aVR, aVL, aVF or V (5 Lead ECG cable)
    Lead I, II, III, aVR, aVL, aVF or V1, V2, V3, V4, V5, V6 (10 Lead ECG cable)
    (Display View : All 12 Lead ECG waves display simultaneously)
    Electrodes Yes
    ECG display Yes
    Message display Yes
    Heart rate display 30 to 300 bpm (±3bpm)
    ECG size 5, 10, 20mm/mV and Auto-gain
    Heart Rate/Arrhythmia Alarm HR, Asystole, VF, VT
    Lead Fault message Yes
    NONINVASIVE PACING
    Waveform Monophasic Truncated Exponential
    Mode Demand and Fixed mode
    Amplitude Accuracy 0 ~ 200mA (±5mA)
    Pulse Width 20ms (±1.5%)
    Pulse Rate 30 ~ 180ppm (±1.5%)
    Refractory Period 340 msec (30 to 80ppm) / 240 msec (90 to 180 ppm)
    SpO2 Pulse Oximetry
    SpO2 monitoring Yes (Nellcor)
    Saturation 70~100% (±3digits)
    Pulse Rate 20~250bpm (±3bpm)
    Perfusion 0.2%
    SpO2 alarm YES (Less than minimum setting rate, Over than maximum setting rate)
    NIBP
    Patient Population Adult, Pediatric, Neonate
    Method Oscillometric
    Control Autometic and manual measurements
    Auto Intervals 1, 3, 5, 10, 15, 30, 60, 120 min
    Displayed Pressures Systolic, Diastolic, Mean mmHg
    Displayer Units Adult : 40 to 260 mmHg
    Pediatric : 40 to 160 mmHg
    Neonate : 40 to 130 mmHg
    Systolic Range Adult : 20 to 200 mmHg
    Pediatric : 20 to 120 mmHg
    Neonate : 20 to 100 mmHg
    Diastolic Range ±3mmHg
    Pressure Trransducer Accuracy Adult : 300 mmHg / Pediatric 300 mmHg / Neonate : 150 mmHg
    Redundant Circuit Overpressure Limit
    PRINTER
    Continuous ECG Strip Real-Time (8 seconds delay)
    Auto Printing Recorder can be configured to print marked event, charge, shock and alarms
    Printing Speed 25mm/s
    Paper 50mm Width / 40mm Diameters
    DATA STORAGE
    Information stored ECG data, Event, Voice
    Internal memory Yes
    Even & ECG record Yes
    Database storage Yes
    External memory card Yes(SD card)
    EXTERNAL LINK
    Data management Yes
    Software Yes(CU-Expert)
    POWER SOURCE
    Type Lithium Polymer (14.8V 3.1Ah – Rechargeable)
    Integral/removable Removable
    Charging method AC adapter / Car Cigar Lighter Jack / both
    AC Power Pack Output : 18V, 6A
    PHYSICAL
    Size 318 X 208 X 355 (W X L X H, mm)
    Weight 5kg (with external paddle)
    DISPLAY
    Type TFT Color
    Size 7 inch Diagonal (152mm X 91mm)
    Resolution 800 X 480 pixels
    AUTOMATIC SELF-TEST
    Power on Self-Test YES
    Run Time Self-Test YES
    Manual Self-Test YES
    Periodic Self-Test Daily / Weekly / Monthly